Measuring Machines

Measuring Machines

Vibration-insulated installation of a SmartScope CNC 670 measurement system with a total weight of 3500 kg.

The manufacturer specifications state that the maximum floor vibration up to a frequency of 15 Hz must be below 1 mg.

This value was greatly exceeded in the vertical direction at our customer, where the value was 5.2 mg. A solution comprising membrane air springs with a vertical natural frequency of 2.5 Hz + a steel platform was initially suggested.

The much simpler, maintenance-free, and more economical isoloc solution was to use our new machine mounting system, the MULTIDAM® 4-axis MD4-16LP/62 (pat.)

Result: there were no problems with vibrations during calibration or upon start-up during normal operation.

Any Questions?